 Brachytherapy for prostate cancer with permanently implanted iodine-125 seeds: development of a protocol dedicated to the dose rate measurement in the vicinity of patients.

Isabelle Clairand, Jean-François Bottollier-Depois, François Trompier1, Viviane Amouroux1, Christian Itié, Céline Texier, Loïc de Carlan, Claudine Dubiau, Didier Franck, Jean-Claude Rosenwald, Bernard Pentsch and Jean-Marc Cosset, IRPA 11, 23-28/05/2004, Madid (Spain).

Summary

  Introduction.  The brachytherapy for prostate cancer with permanently implanted iodine-125 seeds (125I) addresses the problem of radiation safety for people in the vicinity of patients.  The implanted activity on average of 1500 MBq leads to expect significant dose rates outside the patient body (especially during the first year after the implant) which can be assessed by direct measurements on patients.  Such dosimetric data are however very rare in literature, maybe because of the particularity of this type of measurements. Indeed, from a dosimetric point of view, the specificity of 125I lies in the fact that this radionuclide emits low energy photons, below 40 keV.  Consequently, dose rate measurement implies first to know the energy spectrum characteristics and to have detectors whose energy response function is well defined in this energy range.  The goal of this study is to propose a protocol dedicated to the dose rate measurement in the vicinity of patients treated with 125I for prostate cancer.  This protocol is currently applied during a measurements campaign for a representative number of patients.

Material and methods.  Two types of sources (Bebigã and Amershamâ) were studied.  A spectrum analysis was first performed with high-purity germanium detectors for sources placed in free air and under different depths of water in order to check the absence of fluorescence rays and to analyse the spectrum degradation with the distance and the attenuation.  According to the energy spectrum characteristics, the energy response function of three radiameters was studied in a COFRAC-accredited metrology laboratory.  Finally, in order to validate the calibration factors, dosimetric data were measured for the different configurations previously studied from a spectral point of view.  Both spectral and dosimetric analysis were compared with Monte Carlo calculations reproducing the experimental configurations.  Moreover, measurements made in water were compared with calculations performed for a standard anthropomorphic mathematical model for which the prostate was placed under different depths of tissue.

Results.  No significant difference between the two types of sources was shown.  Concerning the spectrum analysis, whatever the configuration, no fluorescence ray was observed and the slight spectrum degradation did not affect the calibration coefficients which were found very stable in this energy range.  The dosimetric analysis allowed to validate, under realistic conditions, the calibration factors for different geometrical configurations and to assess the associated errors.  The consistency between the measurements and the calculations is very good.  Especially, the consistency in terms of spectral and dosimetric range between measurements in water and calculations performed with the anthropomorphic model validates the chosen experimental configurations.  For a prostate located under 4 cm of tissue, the calculated dose rate was found to be 0.13 µGy.h-1.MBq-1 at 25 cm of distance from the patient.  At this same distance, for a prostate located under only 5 mm of tissue, the calculated dose rate was found to be to 0.5 µGy.h-1.MBq-1.

Conclusion.  This study allowed to establish a suitable protocol for the dose rate measurements in the vicinity of patients treated with permanent implant 125I for prostate cancer.


Close

Send to a friend

The information you provide in this page are single use only and will not be saved.
* Required fields

Recipient's email:*  

Sign with your name:* 

Type your email address:*   

Add a message :

Do you want to receive a copy of this email?

Send

Cancel

Close

WP_IMPRIMER_TITLE

WP_IMPRIMER_MESSAGE

Back

Ok