Determination of new European biometric equations for the calibration of in vivo lung counting systems using Livermore phantom

  • La recherche

  • Recherche

01/07/2007

Titre du congrès :IM 2005 - European Workshop on Individual Monitoring of Ionising Radiation
Ville du congrès :Vienne
Date du congrès :11/04/2005
Titre de la  revue : Radiation Protection Dosimetry Vol 125 issue 1-4 Pages : 449-455
Type de document > *Congrès/colloque
Unité de recherche > IRSN/DRPH/SDI/LEDI
Auteurs > DE CARLAN Loïc , FRANCK Didier , PIERRAT Noëlle , PRULHIERE Géraud

New biometric equations used for assessing the thickness of the overlay plate to be added to the physical phantom were determined based on the Computed Tomography (CT) chest images of 33 adult males in order to improve the calibration of in vivo lung counting systems using the Livermore phantom. These equations are specific to systems composed of four germanium detectors with the measured subject in supine position. A comparison with the biometric equations used to date as reference in France was carried out and proved the usefulness of equations directly applicable to the Livermore phantom

Migration content title
En savoir plus
Migration content title
Texte complet
Migration content text
Migration content title
Contact