Photon contribution to ambient dose equivalent H(10) in the wide-spectrum neutron reference fields of the IRSN


Fermer

Authentification

Email :

Mot de passe :


Titre du congrès :NEUDOS 10 
Ville du congrès :Uppsala
Date du congrès :12/06/2006
Titre de la revue : Radiation Protection Dosimetry 126 issue 1-4 Pages 145-150

Type de document > *Congrès/colloque

Mots clés >

Unité de recherche > IRSN/DRPH/SDE/LMDN

Auteurs > ASSELINEAU Bruno, GRESSIER Vincent, LACOSTE Véronique, LANGNER Frank, LEBRETON Léna, LOB S., NOLTE Ralf

Date de publication > 01/08/2007

Résumé

The photon contribution to ambient dose equivalent in several wide-spectrum reference neutrons fields of the Institute for Radiological Protection and Nuclear Safety were measured using a Geiger-Müller counter. For the investigated fields, the ratio of photon to neutron ambient dose equivalent ranged between 0.03 and 0.20. The results show that the Geiger–Müller tube is a versatile instrument for dosimetry in mixed photon-neutron fields if sufficient information for the calculation of corrections is available